These are the medium to large-sized versions of BOD Incubators with similar application of growth & storage of bacterial cultures, gentle incubation & conditioning and Media Preparation.
Walk-In Incubators are ideal for incubation of micro-organisms, eggs, etc to cater to mass-scale production of vaccines & other drug products.
While Humidity is not a parameter to be controlled or measured in a Walk-In Incubator, it is important that there is no loss of humidity in the chamber since it can lead to cracks in the media.
The design team at NEWTRONIC ensure that our Walk-In Incubators are designed to fit into such key requirements perfectly well.
They are supplied at customer-site in a knock-down condition & can pass through standard sized doorways. They are assembled on-site. The installation & commissioning is accompanied with IQ, OQ, PQ protocols & complete documentation.
Features
- Ethernet Communication
- High-speed data transfer
- Significantly faster than serial(RS-485) communication
- Global standard for communication protocol
- No need for a separate PC interface
- Data Operations from any PC over the LAN
- Standby Humidity & Cooling System
- Completely ready-to-use systems
- Automatic switch-over upon failure of primary systems to achieve conditions
- Switch-over also upon detection of HP/LP & Internal fault
- Scheduled & manual change-overs
- Recording of all change-overs in the software
- PLC-Based Control System
- Complete Operations driven through PLC
- No Data Loss: Data Storage of up to 1000 readings on the PLC itself
- At an interval of 30 mins, data up to 20 days is stored on the PLC itself
- Facility to send emails directly from the PLC
- Touch Screen Display
- All operations through a single 4" Color touch screen display
- PLC IP Address setting
- Component Working Hours
- Standby system change over
- Password protection for set values and calibration mode
- E-mail Alert
- When an alarm is triggered, an e-mail is sent to the registered PC or mobile address
- *connection to a mail server is required to use e-mail alert.
- Complete Range of Alarms
- Temperature & Humidity Deviations
- Door Opening
- Water Supply Failure
- Component Failure
- Mains Failure
- Documentation & Training
- IQ, OQ, PQ
- Calibration certificates
- Operations & maintenance manuals
- GA & Electrical drawings
- Material test certificates
- Training to end-users
- Safety
- Temperature Overshoot/Undershoot
- Humidity Overshoot
- Low water-level
- Delay Timer to safeguard the Compressor
- MCB to cut-off over-current supply
- Standby Controlling Sensor upon failure of Primary
- In Walk-in chambers door can be open from inside
- Emergency Bell switch inside the Chamber
- ICDAS Software
- 21 CFR Part 11 Ready
- Single software suitable for up to 20 equipment
- Automatic daily data back-up
- Alarm logs. Alarm acknowledgement with e-signatures
- Audit trails
- MKT Calculations
- Graphical viewing & data trending
- Data export in pdf format
- Touchscreen Door Access System
- Password based access control
- Restricts access to unauthorized persons
- Alarm Transmission
- Mobile alert system
- Time of alarm transmission
- Name of equipment
- Transmitted message(Equipment name, alarm time, type of alarm & limits)
Specifications
| Temperature Range |
10°C to 60°C |
| Temperature Accuracy |
± 0.5°C |
| Temperature Uniformity |
± 1°C |